Before You Start
Only adjust the calibration if a check has shown a deviation greater than 2%. Adjusting the calibration changes how your Cooking Oil Tester (TPM) reads, so only do it when a check tells you to.
If you have not checked the calibration yet, do that first so you know how much to adjust by — see How to check the calibration on your Cooking Oil Tester (TPM).
Steps
Press and hold the H key for 3 seconds to go to the menu settings.
In the menu settings, move up or down using the up and down arrows to select TPM Adjust.
Press the H key to enter the TPM Adjust area.
Using the up and down keys, input the adjustment value. TPM adjustment value = standard value − measurement value.
Press H to save and exit the menu.
Move to Return and press H to return to the measurement screen.
Your calibration has been adjusted. It is worth running another check to confirm the reading is now within 2%.
Still Need Help?
If the tester will not hold its calibration after adjusting, contact the Squizify support team.