How to adjust calibration on Lumity Cooking Oil Tester
Once the calibration has been checked and the deviation is greater the 2% the calibration needs to be adjusted on the Cooking Oil Tester. If you have not checked the calibration you need to complete this step to know how much to adjust the value on the Cooking Oil Tester.
If you are unsure how to check the calibration please see Checking Calibration of the Cooking Oil Tester article.
How to adjust the calibration
Press the H key and hold for 3 seconds to go to the menu settings
In the menu settings, move up or down using the up and down arrows, to select TPM Adjust
Press the H key to enter TPM Adjust area
Using the up and down keys input the adjustment value.
You will know how much to adjust this value once you have completed the Checking Calibration of the Cooking Oil Tester
TPM adjustment value = Standard value - measurement value
Then press H to save and exit the menu
Move to return and press H to return to the measurement screen
Your calibration has been adjusted successfully!
Updated on: 20/03/2024
Thank you!